Power On Status Setup

Background
1. Select Power On
Status
setting
2. Power On Status
Setup appears
Note

Interface

Background
1. Select Interface
setting
64
The Power On Status Setup allows you to either load the
previous settings or the default settings on startup.
From one of the main screens, press
the
ESC
key so that the menu
system at the bottom of the display
has focus.
Go to System and press Enter.
Go to Power On Status Setup and
press Enter.
Use the arrow keys to set Power ON Status Setup. Press
Enter to confirm the setting.
Power On
Status Setup
Power On Status
Pressing ESC before pressing ENTER will exit the Power
On Status Setup.
The remote interface can be set to RS232 or USB.
From one of the main screens, press
the
ESC
key so that the menu
system at the bottom of the display
has focus.
Go to System and press Enter.
Go to Utility and press Enter.
Go to Interface and press Enter.
T3MIL Series User Manual
Enter
Recall Previous Settings,
Load Default
Enter
System
menu icon
Move
Select menu
or setting
System
menu icon
Move
Select menu
or setting