Ergometers with Digital Communication Interface

Revision C
Stress Tests: Ergometers with Remote Start Pulse
Compatible ergometers include Variobike 500, 800, 900/900L, EC1200
V3.52, and Lode Excalibur/Sport.
With these ergometers, the selected protocol fully controls the test
procedure. The ergometer operating controls are inoperative. Only blood
pressure measurements can be taken manually.
127(
The EC1200 will display the version number when turned on. The
following explanations refer to version 3.52 or higher (upgrade, if
possible).
Blood pressure intervals are only valid for Ergoline 900.
The EC1200 with version 3.52 and later only supports directly
connected blood pressure monitors. They are controlled by the stage
time setting. The blood pressure interval will be ignored.
1. Select the ergometer type (Chapter 10, "Communication").
2. Select the test protocol (configuration, Chapter 10,
Mode").
3. Press
. A blood pressure measurement is now taken
automatically, if configured. Wait until the readings are available or
until the first ST measurement is completed.
As trend values are saved at 30-second intervals, it is recommended
that the test subject waits approximately 30 seconds before starting
to pedal.
4. Ask the test subject to start pedalling.
Controls
Holds the stage (must be pressed
5 seconds before the end of a
Alt
stage at the latest).
Stops or cancels the test. Refer to
"End/Cancel a Stress Test"
page 8-7 for more information.
MAC 1200/1200 ST
2012250-021
Advances to the recovery phase
Advances to the next load stage
on
"Stress Test
8-9