10
Specifications
10.1
Reader Specifications
10.1.1
General
Property
Default measurement
technologies
Optional measurement
technologies
Measurement direction
Measurement mode
Plate type compatibility
Wavelength range FI
Wavelength range FP
Wavelength range TRF
Wavelength range LUM
Wavelength range ABS
Temperature control
Temperature uniformity over the
plate area
Air inlet filter
Z-focus range
Shaking modes
Number of optical filters
Number of dichroic mirrors
Instrument connection
VICTOR Nivo – User Manual
Value
Fluorescence (FI)
Luminescence (LUM)
Absorbance, filter-based or spectrometer (ABS
(F) or ABS (S))
Fluorescence polarization (FP)
Time-resolved fluorescence (TRF)
Alpha (ALPHA)
Top, bottom
End Point, Scan (well bottom), Z-Focus Scan
Spectrum scan (ABS (S))
See chapter 10.4 "Plate and Lid Compatibility",
page 73.
230 – 850 nm
400 – 850 nm
230 – 850 nm
230 – 850 nm
220 – 1000 nm
Ambient +3 °C up to +65 °C
37 °C ± 0.5 °C
Spare part order number: HH35900006
0 – 16 mm
Linear, orbital, double orbital
Up to 32
Up to 5 per slide
Ethernet
10 Specifications
65