1. Connect the device to line power by its charging adapter
(WM 28979) and power supply unit and charger.
or
When operating on the portable system: Attach the portable
system to a wall mounting with a charging interface.
or
Connect the device to the vehicle's on-board power supply by
its charging adapter (WM 28979) and 12 V cable.
Result
The device has been connected to the power supply. The device
automatically starts charging the rechargeable battery.

4.3 Connecting the breathing circuit

Hypoventilation resulting from use of additional breathing
system filters!
Using additional breathing system filters (breathing system filter,
bacteria filter or combined breathing system/bacteria filter)
increases the dead space of the overall system. An increased dead
space might lead to hypoventilation. This might expose the
patient to the risk of serious or life-threatening injury.
⇒ Use only approved accessories.
⇒ Pay attention to increased dead space volume when ventilating
at low tidal volumes.