Calibration File Backup - Keysight N9000A Service Manual

Cxa signal analyzer
Hide thumbs Also See for N9000A:
Table of Contents

Advertisement

Post-Repair Procedures
Post-Repair Procedures

Calibration File Backup

Backing up the calibration file before changing the A5 Disk Drive is highly
recommended. To do this you will need to connect a USB mouse and follow the
steps below. This is only available for XSA-Series Software version A.01.55 or
greater. If the file cannot be backed up or the instrument software version is
<A.01.55, all instrument adjustments will need to be run after the drive
replacement.
1. Navigate to the following directory using Windows Explorer:
2. Double-Click BackupAndRestore.exe
3. The following Dialog box will appear:
4. Set the Backup File to the USB drive and save the file.
5. When the A4 Disk Drive is replaced restore this file using the same utility on
While the replacement disk drive has the Windows XP operating system loaded
NOTE
onto it, the license that accompanies it in this kit is for Windows Vista. This is
being done within the scope of our license agreement with Microsoft and is
necessary at this time, as the instrument applications software is not supported or
qualified with any operating system other than Windows XP.
In an effort to eliminate any further confusion for the end user, an additional label
is included in this kit that must be affixed to the rear of the instrument, along with
the Windows Vista license, to explain why the instrument has this Windows Vista
license while running Windows XP.
282
C:\Program Files\Agilent\SignalAnalysis\Physics
the new assembly.
Chapter 14

Advertisement

Table of Contents
loading

Table of Contents