Tests and Adjustments
Adjustments

LO Power Adjustment

The LO power adjustment is used to adjust the power level of the LO signal from the
A6 multiplier board to the A20 mixer brick.
Equipment used for the LO Power Adjustment
Equipment Type
Power meter
Power sensor, 3.5 mm
5/16-inch, open-end torque wrench (set to 10 in-lbs)
Procedure
1. Turn off the network analyzer and remove the power cable and other external cables
from the rear panel.
2. Remove the outer cover. Refer to
3. Place the analyzer on its side to allow access to the bottom of the analyzer.
4. Connect a GPIB cable between the analyzer and the power meter.
5. With a 5/16 inch open-end wrench, disconnect cable W3 at the A20 mixer brick. Refer to
"Bottom Cables, Passive Configuration (Options F06 and F13)" on page 6-16
determine the location of cable W3.
6. Connect the power sensor to the end of cable W3.
7. Reconnect the power and USB cables. Turn on the analyzer.
8. On the System menu, point to Service, Adjustments, and then click LO Power Adjust.
9. Ensure the GPIB settings are correct.
10. Click Begin Adj, and then follow the instructions as they are displayed.
11. Turn off the analyzer, disconnect the power cable, reconnect the semi-rigid cable W3,
and replace the cover.
3-36
Model or
Part Number
E4418B/E4419B
E4413A
N/A
"Removing the Covers" on page
PNA Series Microwave Network Analyzers
Alternate Model
or Part Number
E4418A/E4419A
8485A
N/A
7-6.
to
Service Guide N5230-90014