Performance Test Instructions

Table 22 Reference Wavelength and Power Settings

Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com
Test of the Tunable Laser Module Channel (Slot 0)
If you need to test a back-loadable Tunable Laser module slot, perform the
remaining list items. This test only applies to the Agilent 8164A/B
Lightwave Measurement System.
16Connect the power output of the Tunable Laser module to the input of the
power sensor.
17Switch on the Agilent 8164A/B and wait until is has booted.
18Press Preset, the green hardkey, to set all parameters to their default values.
19Move to [λ], the wavelength of your Tunable Laser module, type 1550.00.
20Move to [λ], the wavelength of your power sensor, type 1550.00.
21Set the [P], the output power, and [Att], the attenuation, to the value given in
Table 22 .
Tunable Laser Module
Agilent 81680A/B, Output 2, High Power
Agilent 81682A/B
Agilent 81682A/B #003
Agilent 81640A/B, Output 2, High Power
22Note the power reading of the power sensor in the test record.
Performance Tests
Attenuation [A
] Power [P]
TT
−10.00 dBm
0 dB
Not applicable
0.00 dBm
−10.00 dBm
0 dB
−10.00 dBm
0 dB