8.4.2

Optional Tests

In this procedure, methods for additional alarm tests are described. These tests are
optional and not needed to ensure safe use of the Vivo 3.
8.4.2.1 High Pressure Alarm
1
Connect the Vivo 3 patient circuit to a test lung and a CPAP device.
2 Set High Pressure alarm to 10 cmH
3 Set the CPAP device treatment pressure to 12 cmH
4 Adjust the Vivo 3 settings as follows:
5 Start the treatment on both the Vivo 3 and the CPAP device.
6 Wait until the High Pressure alarm is activated.
7 Stop the treatment. The test is completed.
Doc. 007232 Y-2
• Ventilation Mode: T
• Target Volume: Off
• IPAP: 10 cmH
O
2
• EPAP: 5 cmH
O.
2
• Breath Rate: 12 bpm
9
9
3
Off
Off
12 bpm
2.0 S
O.
2
O.
2
O.
2
Cleaning and Maintenance 113
Vivo 3 Clinician's Manual