Expiratory Positive Airway Pressure (Epap); Continuous Positive Airway Pressure (Cpap) - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

Property
Setting resolution
Available in modes
4.2.3

Expiratory Positive Airway Pressure (EPAP)

The EPAP setting defines the airway pressure during the expiration phase.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.4

Continuous Positive Airway Pressure (CPAP)

The CPAP setting defines the pressure that will be applied to the airways.
Property
Value range
Default value
Setting resolution
Available in modes
54 Treatment Functions and Settings
Vivo 3 Clinician's Manual
Description
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
• PCV+A
• PSV
• S
• S/T
• T
Description
2 cmH
O to 20 cmH
O.
2
2
For inspiratory pressure settings below 22 cmH
imum value for EPAP is limited to the inspiratory pressure
minus 2 cmH
O.
2
5 cmH
O
2
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
• PCV+A
• PSV
• S
• S/T
• T
Description
4 cmH
O to 20 cmH
O.
2
2
10 cmH
O
2
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
CPAP
O
2
O
2
O, the max-
2
O
2
O
2
O
2
O
2
Doc. 007232 Y-2