Spo 2 Artifact - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
7.4.4
SpO
2
This alarm requires a connected SpO
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Doc. 007232 Y-2
Description
SpO2 Disconnected
Medium
An SpO
if one of the conditions below appears:
• An error signal is received from the sensor
• No signal at all from the sensor is received within 2
Faulty or disconnected sensor.
Normal communication with the sensor is reestablished.
An information message remains until acknowledged by the
user.
The Vivo 3 will continue treatment according to the current
settings.
Artifact
Description
Poor SpO2 Signal
Medium
A poor SpO
low or artifacts are detected by the sensor.
Check the sensor and its placement on the patient.
An OK signal is received from the sensor or the sensor is
disconnected.
An information message remains until acknowledged by the
user.
The Vivo 3 will continue treatment according to the current
settings.
Sensor Failure/Disconnection alarm will be given
2
seconds.
sensor.
2
Signal alarm will be given if perfusion is too
2
Alarms 105
Vivo 3 Clinician's Manual