6.7.1

Using the Attachable Humidifier

WARNING!
Read the section 2.7 Humidification and Heating — Warnings and Precautions, page 26
before using the Vivo 3 with the attachable humidifier.
WARNING!
Risk of Burns
If you remove and open the humidifier just after using the ventilator, ensure that you
don't touch the heater inside the humidifier since it can be very hot.
WARNING!
The humidifier is for single patient use only.
NOTE
The ventilator can also be used with active stand-alone and passive humidifiers
The humidifier is intended to humidify the patient air. It is intended for non-invasive
use only.
About the Attachable Humidifier
The attachable humidifier is intended for non-invasive use only.
The humidifier requires a connection to the Mains power supply to work.
The information in the below table is applicable to the recommended breathing system
configuration, which is Attachable humidifier and Heated circuit.
Property
Humidifier
classification
Rated Flow
Operating Conditions
Gas leakage
Max humidification
output
74 How to Use the Vivo 3
Vivo 3 Clinician's Manual
Value
Category 2 humidifier according to ISO 80601-2-74
20-50 l/min
+15°C to +35°C. Humidity: RH from 15% to 95%.
< 0.2 l/min at 50 cmH
2
> 13 mg/l, tested according to ISO 80601-2-74
O
Doc. 007232 Y-2