Accessories; Accessories/Consumables - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

10

Accessories

The accessories described in this section, together with the medical electric equipment
defined in chapter 3.1 Main Components, page 32, constitute the Vivo 3 medical electric
system.
CAUTION!
Responsibility for System
Accessory equipment connected to the analogue and digital interfaces must be certi-
fied according to the respective IEC standards (e.g. IEC 60950 or IEC 62368 for data
processing equipment and IEC 60601-1 for medical equipment). Furthermore all
configurations must comply with the valid version of the system standard IEC 60601-
1. Anybody who connects additional equipment to the signal input part or signal out-
put part is configuring a medical system, and is therefore responsible for ensuring
that the system complies with the requirements of the valid version of the system
standard IEC 60601-1. If in doubt, consult the technical service department or your
local representative.
A PC, when connected to the ventilator, shall comply with IEC 62368-1, IEC 60950-
1 or IEC 60601-1.
10.1

Accessories/Consumables

Circuit: Single limb 22 mm, disposable
Function: Delivers air to the patient,
applied part
Part No: 005060
Circuit: Single limb 15 mm, disposable
Function: Delivers air to the patient,
applied part
Part No: 006712
130 Accessories
Vivo 3 Clinician's Manual
Doc. 007232 Y-2

Advertisement

Table of Contents
loading

Table of Contents