8.4.2.2 Low Pressure Alarm and the Disconnection Alarm
1. Start the treatment
2. Disconnect the patient circuit.
3. Wait 15 seconds.
⇒The Low Pressure Alarm and/or the Disconnection Alarm will be given.
8.4.2.3 Low Breath Rate Alarm
1
Connect the Vivo 3 to a patient circuit.
2 Set the Low Breath alarm to 10 bpm.
3 Adjust the Vivo 3 settings as follows:
4 Start the treatment on the Vivo 3 device.
5 Wait until the Low Breath Rate alarm is activated.
6 Stop the treatment. The test is completed.
8.4.2.4 High EPAP Alarm
1
Connect the Vivo 3 patient circuit to a test lung and a CPAP device.
2 Set the CPAP device treatment pressure to 10 cmH
3 Adjust the Vivo 3 settings as follows:
Setting
Ventilation Mode
Target Volume
IPAP
EPAP
Breath Rate
Insp. Time
114 Cleaning and Maintenance
Vivo 3 Clinician's Manual
• Ventilation Mode: T
• Target Volume: Off
• IPAP: 10 cmH
O
2
• EPAP: 5 cmH
O.
2
• Breath Rate: 12 bpm
Value
PCV+A
Off
15 cmH
5 cmH
12 bpm
1.5 s
2
O
2
O
2
O.
Doc. 007232 Y-2