Rise Time
Insp. Trigger
4 Start treatment on both the Vivo 3 and the CPAP device.
5 Wait three breaths before the High EPAP alarm.
6 Stop treatment. Test completed.
8.4.2.5 Low EPAP Alarm
1. Make sure Low EPAP alarm is set On and start the treatment.
2. Disconnect the patient circuit from the ventilator.
3. Wait for the duration of 3 breaths before the Low EPAP alarm shall be given.
4. Stop the treatment, the test is completed.
8.4.2.6 High MV Alarm
1. Connect the patient circuit and test lung.
2. Start the treatment.
3. Note the MVe value in the ventilator monitor.
4. Set the High MV Alarm limit below the monitored MVe.
5. Wait approximately 15 seconds before the High MV alarm shall be given.
6. Stop the treatment, the test is completed.
8.4.2.7 Low MV Alarm
1. Connect patient circuit and test lung.
2. Start treatment.
3. Note MVe value in ventilator monitor.
4. Set Low MV Alarm limit above monitored MVe.
5. Wait approximately 15 seconds before the Low MV alarm shall be given.
6. Stop treatment, test completed.
8.5

Repair

Repairs of the Vivo 3 must be carried out by authorized personnel only, in accordance
with instructions from Breas Medical.
Authorised service workshops can order the Vivo 3 service manual that contains all
technical documentation required for the maintenance of the Vivo 3.