3 If you will use the Vivo 3 without the humidifier, install the air bypass unit in place
of the humidifier.
Fill the Humidifier
Duration of operation between humidifier refills
Humidifier level (5): 12 hours. At lower settings, the duration will be longer.
CAUTION!
• The water chamber must be detached from the Vivo 3 when filling water into the
water chamber.
• Use only distilled or sterilized water or boiled, chilled tap water in the humidifier
water chamber. This is to reduce bacteria and mineral deposits.
• Do not fill the water chamber with hot water.
• Do not overfill the water chamber. The water chamber has a capacity of 350 ml
and the maximum filling level is indicated on the chamber.
• After using the ventilator, wait one minute before you open the water chamber
since it can get hot under certain conditions (for example if the humidifier runs
out of water).
1
Detach the water chamber, see above.
2 Make sure the water meets the quality requirements. It shall be either:
• Distilled
• Sterilized
• Boiled and chilled tap water.
3 Hold the humidifier with the air
path openings (a) up and fill water
into either of the air path openings.
Make sure not to overfill (b).
76 How to Use the Vivo 3
Vivo 3 Clinician's Manual
a
a
b
Doc. 007232 Y-2