Breas Vivo 3 Clinician Manual page 92

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

Information Message Audio Signal
Short signal every 5 seconds
Alarm Sound Pressure Level
The alarm sound pressure level is adjustable within 55–80 dBA.
Alarm Message on the Screen
The name of the active alarm is displayed on the screen.
• High priority alarms
Red highlight color.
• Medium priority alarms
Yellow highlight color.
For detailed information about specific alarms, see 7.3 Physiological Alarms, page 95
and 7.4 Technical Alarms, page 104.
Display of Multiple Alarms
If several alarms are active, active high priority alarms have precedence over medium
priority alarms: All high priority alarm conditions must be resolved before any medium
alarms are displayed.
If several alarms of the same priority are active at the same time, the alarm descriptions
are looped in the display.
A ">>" symbol indicates that more alarms are to be displayed in the loop.
Alarm LED signal
• High priority alarms
Red light, flashing quickly (0.5 s.
interval).
• Medium priority alarms
Yellow light, flashing slowly (2 s.
interval).
92 Alarms
Vivo 3 Clinician's Manual
Doc. 007232 Y-2

Advertisement

Table of Contents
loading

Table of Contents