Connect The Heated Patient Circuit; Disconnect The Heated Patient Circuit From The Ventilator; Power Up The Vivo 3 - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

1
Check that the circuit is clean and undamaged.
2 Connect the tube to the air outlet.
If having a heated circuit, make sure to connect the end with the heating plug to the
to the air outlet. Refer to 5.4.1 Connect the Heated Patient Circuit, page 66 and 5.4.2 Dis-
connect the Heated Patient Circuit from the Ventilator, page 66.
3 Check whether the patient interface has an integrated leakage port.
• If yes, connect the patient circuit to the patient interface.
• If no, first connect a leakage valve to the patient circuit, then connect it to
A bacterial filter can be attached on the ventilator outlet connector.
5.4.1

Connect the Heated Patient Circuit

1
Check the patient circuit for damages.
2 Connect the ventilator cuff to the
air outlet of the ventilator. A click-
ing sound is heard when the latches
are fitted correctly.
3 If applicable, perform a pre-use test on the ventilator. For specific information, see
the User Manual for the ventilator.
4 Connect the patient interface cuff to the patient interface.
5 On the ventilator, activate the circuit heating. .
5.4.2

Disconnect the Heated Patient Circuit from the Ventilator

1
Disconnect from the ventilator:
Press the latches and pull the circuit off from the ventilator. Do not pull by the tube.
2 Disconnect from the patient interface:
Hold by the cuff and pull the circuit off from the patient interface. Do not pull by
the tube.
5.5

Power up the Vivo 3

This function check may be performed as a pre-use check or whenever the function of
the Vivo 3 needs to be checked.
66 Prepare the Vivo 3 for Use
Vivo 3 Clinician's Manual
the patient interface.
Doc. 007232 Y-2