Mve (Minute Volume, expiratory) is calculated as the Tidal Volume multiplied with the
Total Breath Rate.
• Vte
Vte (Tidal volume, expiratory) is the expired tidal volume for each breath.
• % in TgV
Percentage of breaths within the target volume window.
• Rise Time
Rise Time displays the duration of the pressure increase during the inspiration phase.
• SpO2
SpO
ured by the SpO
This value is only displayed if the SpO
Displayed ranges: According to the manufacturer's specifications.
• Pulse Rate
Pulse Rate displays the patients pulse rate as measured by the SpO2 module.
This value is only displayed if the SpO
Displayed ranges: According to the manufacturer's specifications.
6.6.2

Treatment Values Monitored by External Equipment

Monitoring of Expiratory CO
For monitoring expiratory CO
patient circuit.
The device shall comply with ISO 80601-2-55.
Monitoring of Oxygen Saturation
For monitoring the patient air oxygen saturation, an external monitoring device shall be
connected to the patient circuit.
The device shall comply with ISO 80601-2-55 and have a high oxygen level alarm.
6.7

Using Accessories

This section describes how to use accessories provided by Breas Medical.
Doc. 007232 Y-2
(Saturation of Peripheral Oxygen) displays the patient's oxygen saturation. as meas-
2
module.
2
2
module is connected.
2
module is connected.
2
2
, an external monitoring device shall be connected to the
How to Use the Vivo 3 73