Maximum Inspiratory Time (Max. Insp. Time)

The Maximum Inspiratory Time setting defines a maximum length for each inspiration.
If the Maximum Inspiratory Time is set to Off, the length of the inspiration and/or
minimum inspiratory time is dependent on the set Expiratory Trigger.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.9

Inspiratory Trigger (Insp.Trigger)

The inspiratory trigger setting defines the patient's effort required to initiate a ventilator
assisted breath.
When the patient starts a breath, an increasing flow is created in the patient circuit. If
the patient's effort reaches the set inspiratory trigger level an inspiration is initiated. If
the patient cannot trigger a breath, the ventilator will deliver breaths according to the
set Breath Rate.
The Assisted breath modes in PCV are turned off if the inspiratory trigger is set to Off.
Property
Value range
Default value
56 Treatment Functions and Settings
Vivo 3 Clinician's Manual
Description
0.1 s
• PSV
• S
• S/T
Description
• Off
• 0.3 s to 3 s
Off
0.1 s
• PSV
• S
• S/T
Description
Auto, 1 to 9
Auto