4.2.16 Ramp Down
The ramp Down setting defines a ramp time for decreasing the airway pressures.
At the end of the Ramp Down time, IPAP will end 2 cmH
EPAP.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.17 Ramp Pressure
The ramp pressure setting is used for defining the start pressure when Ramp Up is
used:
IPAP shall start 2 cmH
Pressure is lower than EPAP
CPAP mode:
Property
Value range
Default value
60 Treatment Functions and Settings
Vivo 3 Clinician's Manual
Description
• Off
• 10 min to 60 min
Off
10 min
• PCV+A
• PSV
• S
• S/T
• T
• CPAP
O above Ramp Pressure, or 2 cmH
2
Description
• 2 cmH
O to the set value for EPAP (PSV and PCV+A
2
modes)
• 2 cmH
O to the set value for CPAP (CPAP mode)
2
• The set value for EPAP (PSV and PCV+A modes)
• The set value for CPAP (CPAP mode)
O above the initially set
2
O above EPAP if Ramp
2
Doc. 007232 Y-2