Recommended Separation Distances Between External Power; Delivery Settings - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

• These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
9.7.4

Recommended separation distances between external power

conductors and the Vivo 3
Rated maximum current in
conductor (A)
1
10
30
For conductors rated at a maximum current not listed above, the recommended separa-
tion distance d in meters (m) can be estimated using the equation d=I/2πH, where I is
the maximum current rating of the conductor in amperes (A) according to the transmit-
ter manufacturer; H is the Vivo 3 immunity compliance level to electromagnetic fields
in the 50-60 Hz frequency span (30 A/m).
9.8

Delivery Settings

Alarm
Low Pressure
High Pressure
High Pressure
High Minute Volume
Low Breath Rate
High Breath Rate
Apnea
Low Pulse Rate
High Pulse Rate
Low SpO2
High SpO2
Disconnection (of mask)
Rebreathing (blocked mask leakage)
Low EPAP
High EPAP
Alarm Sound Level
128 Technical Specifications
Vivo 3 Clinician's Manual
Separation distance (m)
50-60 Hz: d= I/2πH= I/188
0.005
0.05
0.16
Default Setting
10 cmH
25 cmH
Off
Off
Off
Off
Off
Off
Off
90%
Off
On
On
Off
Off
5
O
2
O
2
Doc. 007232 Y-2