Double-Pulse Test Preparation

Connect the 10076C High Voltage Probe

Take extra care when inserting the 10076C High Voltage Probe in the probe
socket on the DUT module or in the Deskew socket. While it is not difficult, you
must ensure the probe tip is fully inserted in the socket center pin.
The probe tip must be
inserted all the way into
the tip socket.
High Voltage
Oscilloscope
Probe.
Probe Tip
Socket
Click on the following screen for a short movie showing how to install the High
Voltage probe. Note, after the movie starts, right click and select Full Screen
Multimedia to enlarge the video.
40
When inserting the 10076C High Voltage probe into the socket on
the DUT module or the Deskew socket, make certain that the probe
tip goes into the socket center pin. The shoulder of the probe
should rest against the top of the socket.
Probe shoulder
is against top
of the socket
Keysight Double-Pulse Test Fixture Installation and Use Guide
Connect Oscilloscope Probes to Oscilloscope
X
Probe shoulder
is not against
top of the
socket