Use key 1 and key 2 to enter the maximum load of the loadcell(s). Key 1 is used for
changing the number (1-9), key 2 is used for changing the position of the cursor. And
press key 3 >2 sec.to set the maximum load.
Up
The following screen is visible:
In tcL 2 you can set the loadcell sensitivity for loadcell 1. This information can be found on
the datasheet delivered with the loadcell. Press key 2 <2 sec. to set the loadcell
sensitivity.
The following screen is visible:
Use key 1 and key 2 to enter the sensitivity of loadcell 1. Key 1 is used for
changing the number (1-9), key 2 is used for changing the position of the
cursor. Sensitivity has to be filled in as 0.00000 mV/V. Press key 3 >2 sec. to set the
sensitivity of the loadcell.