MANUAL SGM800
Example Max Step:
If the settings are:
Step size = 1, Range = 100 and Max. Step = 50,
the table on the right shows the
accompanying step size with which the
weigher values reduces within the displayed
ranges.
When the indicator is set certified, the
maximum preset tare is equal to the first
level of the autorange. In this example the
preset tare is valid to 100.
The following screen is visible:
In rng 3 you set the auto range reset option. Choose between:
of: Multi Range = the highest shown step size will be reset after the signal has been
lower or equal to zero
on: Mulit Interval = the highest shown step size will be reset after the signal reaches
the previous range.
To change the auto range reset option press key 2 < 2 sec.
The following screen is visible:
Use key 1 change the reset option. Confirm by pressing key 3 for >2 sec.
Up
Down Confirm
Displayed range
0-100
100-200
200-500
500-1000
1000-2000
2000-5000+
Step size
1
2
5
10
20
50
67