ETC PENKO SGM800 Manual page 77

Table of Contents

Advertisement

MANUAL SGM800
First calibrate the zero point (CP1). Make sure the weigher is unloaded and press key 3
>2 sec.
The indicator now shows CP2 to calibrate the gain point (CP2).
And will automatically jump to:
Use key 1 and key 2 to enter the reference value. Key 1 is used for changing the
number (1-9), key 2 is used for changing the position of the cursor. Load the weigher
with the reference value and press key 3 >2 sec.
Up
Down
Left
Right
Confirm
Done successfully the following screen is visible:
In CAL 2 you can check the weigher information. You can check out the, actual
weigher value, actual weigher value x10 and the actual ADC value.
Press key 2 <2 sec. to check out the weigher information.
The following screen is visible:
77