MANUAL SGM800
The following screen is visible:
Use key 1 and 2 to change the value. Key 1 is for changing the number (1-9), key
2 is for changing the position of the cursor. Confirm by pressing key 3 for >2 sec.
Increase Decrease Left
The following screen is visible:
In Ind 4 you set the Digital overall filter. Set the overall filter to effect all indicator
signals used in the device. 0dB means no effect and –50dB is the strongest damping.
Choose between -: 0dB, 1: -6dB, 2: -12dB, 3: -24dB, 4: -30dB,
5: -36dB, 6: -42dB and 7: –50dB.
To change the filter press key 2 < 2 sec.
The following screen is visible:
Use key 1 change the filter. Confirm by pressing key 3 for >2 sec.
Up
To prevent a loss of information or accuracy, don't set
the overall filter higher than 24dB. When no accuracy is
needed, a higher filter setting is allowed
to enable extreme filtering.
Down Confirm
Right
Confirm