MANUAL SGM800
2.1.12 Menu -dSF-
dSF
Digital filter settings
In this menu, the digital filter settings can be set. This filter is a 2
effects all signals up to and including the cutoff frequency. Options are:
dSF 1
dSF 2
dSF 3
Press key 2 <2 sec to enter the digital filter settings.
The following screen is visible:
In dSF 1 you can set the filter type.
Choose between None, Dynamic and Static. Dynamic application = used when the
weighing signal is constantly changing. Static application = used when the weighing
signal is stable. Press key 2 <2 sec. to change the filter band.
The following screen is visible:
Use key 1 change the filter type. Confirm by pressing key 3 for >2 sec.
Up
Filter type
Cutoff frequency
Moving average cutoff frequency
Down Confirm
nd
order filter. The filter
71