SA Mode Settings
Select SA Mode before making any setting in this chapter.
Because there is no calibration, these settings can be made in any order.
How to select SA Mode
Press Mode_.
Then SA.
Frequency Range
The X-axis frequency range determines the frequencies that are measured for each
sweep. The default Start frequency is 0 Hz. However, the Start frequency can be set
as low as -100 MHz. The internal LO of the FieldFox can be seen at 0 Hz, which will
mask signals that may be present.
NOTE
Although the start frequency can be set as low as -100 MHz, amplitude accuracy is
specified above 2 MHz. Below 2 MHz, frequency accuracy is maintained, but
amplitude accuracy is degraded.
68
o
FieldFox User's Guide