Keysight N9912A User Manual page 95

Fieldfox analyzer
Hide thumbs Also See for N9912A:
Table of Contents

Advertisement

The following two Channel Power levels are displayed:
 Channel Power - measures total power over the specified Integrated BW.
 Power Spectral Density – same measured power, but mathematically normalized
How to select Channel Power
 Press Measure 1
 Then
 Then
When Channel Power is selected, the following settings are maintained from a
previous measurement: Center Frequency, Preamp ON|OFF, and RF Attenuation.
Integration Bandwidth
When Channel Power is selected, vertical posts appear on the display to mark the
current Integration Bandwidth setting. The displayed Channel Power and Power
Spectral Density values are measured and calculated over the specified Integration
Bandwidth.
By default, the displayed frequency span is automatically coupled to the Integration
Bandwidth. As you change the Integration Bandwidth, the frequency span is
adjusted so that the vertical posts appear to NOT move. However, when you
manually change the frequency span, the Integration Bandwidth is no longer coupled
to the frequency span.
When a Radio Standard is selected, the appropriate Integration Bandwidth is set
automatically. Learn more about Radio Standards on page 70.
To change Integration Bandwidth:
 Press Meas Setup 4
 Then
SA (Spectrum Analyzer) Mode
Channel Power measurement; Chan 190 Downlink; GSM850 Radio Standard
to a 1 Hz bandwidth using the same algorithm as Noise Marker. Learn more about
the Noise Marker on page 90.
Channel Measurements
Channel Power
Integrating BW
95