How to perform O,S,L Cal
Before performing an O,S,L Cal, set the Measurement type, Frequency Range, Power
Level, and Resolution of the measurement. The calibration will remain highly
accurate until these settings change, or the physical connection to the FieldFox
changes.
 Disconnect the DUT from the FieldFox.
 If a jumper cable or adapter is required to connect the DUT to the FieldFox, then
 In CAT or NA Mode, press Cal 5.
 Then (if necessary) press
 Then (if necessary) press
 Then using the ▲|▼ arrows or the rotary knob, select the connector type and
 If there is more than one Cal Kit for the selected connector type, press
NOTE
Calibration kit definitions are unique to each connector type and connector sex. They
must be correctly applied to obtain the most accurate calibration.
 Then press
 Press Open, Short, and
 Connect the corresponding standard at the point where the DUT will be
 Measure each of the three standards. Standards can be measured again at any
 Press

Normalize

Normalize is used to calibrate a 2-port Insertion Loss measurement (CAT Mode) and
a S21 Transmission measurement (NA Mode). One of these measurements must be
selected before performing a Normalize Cal.
NOTE
A Normalization Cal is REQUIRED to make accurate 2-port Transmission or Insertion
Loss measurements.
How to perform a Normalization
 Press Cal 5.
Calibration for CAT, NA, and VVM Modes
is shown on the screen when an O,S,L Cal is correcting the
connect those components to the FieldFox RF Out connector NOW. The effects of
those components will be measured and removed during the calibration, and only
the effects of the DUT will be displayed in the measurement results. These should
be high-quality components!
gender of your DUT port to be measured and press Enter. This will determine the
type and gender of Cal Kit to use. For example, if a Type-N Male connector is used
on the DUT port to be measured, then the Type-N Male standards (OPEN, SHORT,
AND LOAD) will ALL be connected in place of the DUT during the calibration.
select the correct Cal Kit to use.
Start
Cal.
connected, then press Measure.
time if you feel a standard was not connected securely.
to complete the O,S,L calibration.
Finish
should appear.
Cal Type [Normalize]
to choose
Cal Type
O,S,L
Connector
Type.
in any order.
Load
.
Cal Kit
and