In the following table, Data/Mem shows both pairs of connectors to have the same 2
dB difference. However, the second pair of connectors have much better S11
performance (–50 and –52) and the relative significance is shown in the Data-Mem
values.
Data Analysis Features
Values to compare
Data/Mem
–10 dB and –12
2 dB
dB
–50 dB and –52
2 dB
dB
Data-Mem
for S-parameters
for absolute power levels
with 20 *log
with 10*log relationship
relationship
-24 dB
-64 dB
Data-Mem
-14 dB
-54 dB