Set Frequency/Span; Radio Standard - Keysight N9912A User Manual

Fieldfox analyzer
Hide thumbs Also See for N9912A:
Table of Contents

Advertisement

CAUTION
Frequency/ Span / Step
Valid CPM measurements can be made over the entire frequency range of the
FieldFox. Although it is possible to enter a center frequency below the FieldFox
minimum frequency, the LO feed-thru of the internal SA hardware will invalidate
these Built-in Power Meter measurements.
The measurement span (also known as Channel Width and Integration BW) is the
frequency range, half above and half below the center frequency, over which power
is detected and displayed. The maximum measurement span is 100 MHz.
A Radio Standard can also be used to set the center frequency and span. Learn more
about Radio Standards on page 70.
How to enter center frequency, step size, and span
The default frequency is 50 MHz.
 Press Freq/Dist or Meas Setup 4 .
 Then Frequency.
 Then
 Then type a value using the numeric keypad. Then press a multiplier key. Learn
 Optionally, use the ▲|▼ arrows to quickly change Center frequency. The size of
 Optionally press

Radio Standard

A Radio Standard can used to integrate power within specific radio channels. When a
Radio Standard is selected, the Span setting is automatically coupled to the defined
Channel Width for that standard. (Width is the same for all channels within a
standard.) Also, Unit is set to Chan and CF is set to the nearest specific Channel
defined in the standard.
Learn all about Radio Standards, including Importing and Deleting Custom Radio
Standards, beginning on page 70.
How to select a Radio Standard
 Press Measure 1 OR Freq/Dist
 Then
 Then select a standard using the ▲|▼ arrows or rotary knob and press Enter.
Channel Power Meter Mode - Option 311
Please observe the RF IN port damage level.
+27 dBm (320 mW) average
+33 dBm peak (2 W) <10μs
+50 V
Center
about multiplier abbreviations on page 19. The current frequency (f) is shown on
the FieldFox screen.
the frequency step can also be changed. Press
value using the numeric keypad, then select a multiplier key.
Span (Ch Wid)
press a multiplier key. The default span is 2 MHz. The maximum measurement
span is 100 MHz. The current span setting is shown as the dividend in the digital
display. For example: -10 dBm / 1.00 MHz
Radio Standard
then type a step size
CF Step
Type a value using the numeric keypad. Then
115

Advertisement

Table of Contents
loading

Table of Contents