Attenuation

The Attenuation setting controls the power level into the FieldFox.
To compare CPM measurements to SA Channel Power measurements, use the same
attenuation settings. Built-in Power Meter Mode defaults to 30 dB attenuation, while
SA mode defaults to 10 dB attention. In CPM Mode, the attenuation setting is NOT
annotated on the screen.
How to set Attenuation
 Press Scale/Amptd
 Then
 Then
 Then choose from either 30 dB or 10 dB
Note:
The following CPM settings are identical to the standard Power Meter settings. To
learn about these settings, please refer to the following pages:
Set Scale ............................................................. 119
Set Relative and Offset Power ............................ 120
Set Display Units ................................................ 121
Set Averaging ..................................................... 121
Single/Continuous .............................................. 122
Set Limits ............................................................ 122