Keysight N9912A User Manual page 47

Fieldfox analyzer
Hide thumbs Also See for N9912A:
Table of Contents

Advertisement

NOTE
Measurement speed specifications do NOT apply in Temperature Control Mode.
Learn more on page 14.
How to set Sweep Time
 Press Sweep 3.
 Then
 Enter a value using the numeric keypad.
 Press a multiplier key. Learn about multiplier abbreviations on page 19.
Output Power
Set the power level out of the FieldFox to High, Low, or manually set power level to a
value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss devices
to place the signal farther from the noise floor. However, for devices that are sensitive
to high power levels such as amplifiers, use the Low power setting.
For best measurement accuracy, use the Manual power setting at -15 dBm. After
calibration, the power level can be decreased for amplifiers, or increased for higher
dynamic range.
Caution
Power Level settings in this mode will NOT change Power Level settings in other
modes. To help prevent damage to your DUT, use caution when changing modes with
your DUT connected to the FieldFox test ports.
How to set Output Power
Output power is NOT leveled with any of the power settings:
 Press Meas Setup 4.
 Then
System Impedance (Z0)
To accurately view data presented in Smith Chart format, first set the System
Impedance.
Learn how to select Smith Chart format on page 41.
Learn how to make 75Ω measurements at the FieldFox Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/FieldFox.htm
How to set System Impedance
 Press Meas Setup 4.
NA (Network Analyzer) Mode
Min Swp
Time.
Output Power
High (Default setting) Sets output power to the maximum achievable power at
o
all displayed frequencies. Please see the FieldFox Specifications (page 201) for
expected power levels.
Low Sets output power to approximately –27 dBm,
o
Man Set output power to an arbitrary value,
o
Then press
Nominal Power
o
Then enter a value using the numeric keypad, the ▲|▼ arrows, or the rotary
o
knob.
Press Enter.