1-Port Cable Loss Measurements - Keysight N9912A User Manual

Fieldfox analyzer
Hide thumbs Also See for N9912A:
Table of Contents

Advertisement

When measuring a LOAD, very little power is reflected and values of 40 dB to 60 dB
are displayed.
The minus sign is usually ignored when conveying return loss. For example, a
component is said to have 18 dB return loss, rather than –18 dB.
How to measure Return Loss
 Connect the cable or any adapter used to connect the device under test (DUT).
 Select Preset then
 Select Mode then
 Then
 Press Freq/Dist and enter
 Press Meas Setup 4 then
 Disconnect the cable or DUT and press Cal 5 then follow the calibration
 Reconnect the cable or DUT.
 The return loss trace is displayed on the FieldFox screen.

1-Port Cable Loss Measurements

While all cables have inherent loss, weather and time will deteriorate cables and
cause even more energy to be absorbed by the cable. This makes less power
available to be transmitted.
A deteriorated cable is not usually apparent in a Distance to Fault measurement,
where more obvious and dramatic problems are identified. A Cable Loss
measurement is necessary to measure the accumulated losses throughout the length
of the cable.
A 2-port Insertion Loss measurement is usually more accurate than a 1-port Cable
Loss measurement. However, to perform a 2-port Insertion Loss measurement, both
ends of the cable must be connected to the FieldFox, and the FieldFox must have
option 110 installed.
NOTE
In high-loss conditions, a Cable Loss measurement becomes 'noisy' as the test signal
becomes indistinguishable in the FieldFox noise floor. This can occur when
measuring a very long cable and using relatively high measurement frequencies. To
help with this condition, use High Power (page 25) and Averaging. (page 23).
How to make a 1-port Cable Loss Measurement
1. Press Preset then Preset.
2. Then
3. Connect the cable to be tested.
4. Press Freq/Dist and enter
5. Press Sweep 3 then
6. Remove the cable to be tested.
CAT (Cable and Antenna Test) Mode
Preset
CAT
(Default measurement).
Return Loss
calibrating.
prompts.
then
More
Cable Loss (1-Port)
measurement.
trace is visible on the screen. The amount of time that is required increases with
longer cable lengths. Learn more in the Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/FieldFox.htm
Returns the FieldFox to known settings.
(Cable and Antenna Test)
and
frequency values of the measurement.
Start
Stop
to make appropriate settings before
Settings
.
and
frequency values of the
Start
Stop
Time. Increase the Sweep Time until a stable
Min Swp
27