JCM GLOBAL UBA Pro Series Operation And Maintenance Manual page 76

Universal banknote acceptor
Table of Contents

Advertisement

Section 6
9. Confirm that the Sensor Calibration Program
Screen appears (Figure 6-25).
10. Click the "
Sensor Calibration
6-25 a).
Figure 6-25 Sensor Calibration 3
a
Figure 6-25 Sensor Calibration 3
11. Click the "
"
Start
begin the Sensor Calibration "Without Paper".
NOTE: If the Start button is grayed out
(off), check the USB cable connection.
Figure 6-26 Sensor Calibration 4
Figure 6-26 Sensor Calibration 4
12. 1st "
Calibrate DA without paper
"
Calibrate time without paper
matically run.
Figure 6-27 Sensor Calibration 5
= Done
= In-operation
In-operation Item
Progress Bar
Figure 6-27 Sensor Calibration 5
P/N 960-000211R_Rev. A
UBA™ Pro Series Universal Banknote Acceptor
"
button (Figure
button (Figure 6-26 a) to
a
" through 3rd
" items will auto-
= Error
Reset to restart Calibration
when an error occurs.
13. The Sensor Calibration will pause before the
"With Paper" calibration.
14. Open the UBA Pro's Cover (Figure 6-28 a).
15. Place the Reference Paper (Figure 6-28 b).
(Refer to "Placing Reference Paper" on page 6-5
for detailed instructions.)
Figure 6-28 Sensor Calibration 6
b
Figure 6-28 Sensor Calibration 6
16. Close the UBA Pro's Cover until it "clicks" into
place.
17. Click the "
"
OK
begin the Sensor Calibration "
Figure 6-29 Sensor Calibration 7
Figure 6-29 Sensor Calibration 7
18. 4th "
Calibrate gain with paper
brate bar DA with paper
matically run.
Figure 6-30 Sensor Calibration 8
= Done
= In-operation
In-operation Item
Progress Bar
Figure 6-30 Sensor Calibration 8
© 2020, JAPAN CASH MACHINE CO., LTD.
6 - 6
Calibration and Testing
a
button (Figure 6-29 a) to
".
With Paper
a
" through 5th "
" calibration will auto-
= Error
Reset to restart Calibration
when an error occurs.
Cali-