Replacing The Valve Cartridge - Waters ACQUITY CM-30S Overview And Maintenance Manual

30-cm single zone column manager
Table of Contents

Advertisement

Replacing the valve cartridge

This procedure is only applicable to CM-30S module installed in a UPC
Note:
system.
Before beginning the replacement procedure, ensure that you have the correct
part for your system type.
Warning:
including inhalation, observe Good Laboratory Practice when you
handle them. See the Material Safety Data Sheets for the solvents you
use.
Notice:
chemical-resistant, powder-free gloves when replacing the CM-30S
valve cartridge.
Required materials
Gloves: clean, powder-free, chemical-resistant
Valve cartridge
1/4-inch open-end wrench
2-mm Allen wrench (startup kit)
To replace the CM-30S valve cartridge:
1.
Stop any running mobile phase before replacing the CM-30S valve
cartridge.
2.
In the instrument console, select Column Manager from the system tree.
3.
In the column manager information window, click Interactive Display.
4.
In the column manager interactive display dialog box, click Control.
5.
Ensure the switching-valve is set to Waste.
To change the setting to Waste, click System and select Waste.
Tip:
To avoid the harmful effects of personal contact with solvents,
To prevent contaminating system components, wear clean,

Replacing the valve cartridge

October 7, 2014, 715004178 Rev. B