Operating Instructions SIROLaser Advance
+
+
63 74 198 D3545
D3545.201.06.03.02 05.2014
8.6
Replacing the batteries of the wireless foot
control
The wireless foot control is powered by two (2) AAA batteries
(commercially available).
When the battery is empty, an error code is output, see "Error messages
and warnings". In this case select the finger switch in the set-up submenu
"Activation device" [ ➙ 46] for further operation of the SIROLaser
Advance.
The batteries can be changed by the user.
The housing of the wireless foot control must be opened to change the
battery. Touch a grounded metal part before opening the housing to
prevent damage to the PC board due to electrostatic discharge.
CAUTION
Prior to changing the batteries, switch the SIROLaser Advance off at the
main switch. This prevents accidental triggering.
Removing and replacing the batteries
1. Remove the screws from the bottom of the foot control.
2. Remove the cover and open the battery compartment.
3. Pull the battery holder out of the battery compartment and replace the
batteries with new ones. Be careful to insert it with the correct polarity
(minus pole facing spring).
Assembling the foot control
1. Place the battery holder back again in the battery compartment.
2. Close the battery compartment with the cover.
3. Screw tight the screws at the bottom of the foot control.
NOTICE
After changing the batteries, start the SIROLaser Advance and check
the complete functionality of the foot control. In case the finger switch
was selected as preliminary activation device it is necessary to re-select
the wireless foot control. It is not necessary to reregister the foot control
again at the SIROLaser Advance after changing batteries.
8.7

Replacement of parts subject to wear and tear

Check the following parts subject to wear and tear and replace where
applicable:
With foot control and battery
● Optical fibers (change after 10 treatments/sterilizations)
● Single-use tips (change after every treatment)
● Handpiece sleeve incl. plastic body and silicone switch coat (change
after 400 treatments/sterilizations)
● Bending tool for single-use tips (change after 100 treatments/
sterilizations)
Maintenance and service

Replacing the batteries of the wireless foot control

69