DD+DIS155.11E
(3)
Create a new exam "System
Diagnosis - Flatfield".
(4)
Scan the image.
IMPORTANT:
By selection of examination type <System Diagnosis Flatfield>,
the correct settings for image processing of the flatfield are used.
Any other examination type gives a result which cannot be used for image evaluation.
(5)
Check the image for the
following artifacts:
NOTE:
The slow scan direction is parallel
to the long side of the image plate.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
03-2013
Controls, Connections, and Setup Procedures
Stripes in fast scan or
slow scan direction
Large area
inhomogeneities
Unacceptable number
of white dots
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Figure 63
Figure 64
Slow Scan
Direction
Image
Plate
Figure 65
Chapter 1 / Page 44 of 47
Agfa Company Confidential
System
Diagnosis
Genrad
Flat
Field
200
install_poster_14a.cdr
install_poster_15.cdr
Monitor
515101aa.cdr