AGFA CR Series Service Manual page 32

Digitizer
Table of Contents

Advertisement

DD+DIS238.06E
Replacing batteries:
WARNING:
Battery can explode, causing chemical burns.
INSTRUCTION:
Check that batteries are inserted with correct polarity.
Only use batteries of the same type or an equivalent type as specified by the
manufacturer.
Dispose of empty batteries in compliance with the specifications of the manufacturer.
When removing lithium batteries from the equipment take appropriate measures to
avoid short circuit of the battery:
Either use tape to cover the two poles of the battery or put the battery back in its
original packing and secure the packing by tape.
Performing the electrical test according to national regulations:
WARNING:
Improper ground connections inside the device or too high leakage current may lead to
electric shocks.
After any work at the power supply or at any component connected to mains voltage
inform the responsible organization
according to national regulations.
If specific national regulations do not exist: It is recommended to perform the
electrical test according to IEC 62353.
Make sure, that all grounding connections to metallic covers and all grounding
connections inside the device are present.
NOTE:
On MedNet, GSO Library path "General Info => Agfa HealthCare => Publications =>
Service Manual" a form "IEC 62353:2007 Test Documentation" for CR Digitizers is
available, to be handed out to the responsible organization
1
Responsible Organization:
Entity accountable for the use and maintenance of a medical equipment or a medical equipment system. The
accountable entity can be, for example, a hospital or an individual clinician.
Edition 1, Revision 3
07-2009
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Generic Safety Directions for HealthCare Imaging Products
1
about the necessity of the electrical test
DOCUMENT CONTROL NOTE:
Generic Safety Directions
(Intranet Link
/
Extranet
Page 24 of
Agfa Company Confidential
Link).