AGFA CR Series Service Manual page 199

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
1.2
LEDs on the PMI Board
LED
LD1
LD3
LD5
LD7
LD9
LD10
LD11
LD12
LD13
LD14
LD15
LD16
* Default during operation
** On only if there is ethernet traffic
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Electrical and mechanical Codes, Fuses, LEDs
Status Indicator
Green
Constant
Green
Constant
Green
Constant
Green
Constant
Green
Constant
Green
Constant
Green
Constant
Green
Constant
Green
Constant
Red
Constant
Yellow
Constant
Green
Constant
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Meaning
ON*
5 V for USB ok
OFF
5 V for USB missng
ON*
24 V for Galvo OK
OFF
24 V for Galvo missing
ON*
3,3 V for on board components OK
OFF
3,3 V for on board components missing
ON*
5 V for on board components OK
OFF
5 V for on board components missing
ON*
Supply voltages for Optics OK
OFF
Supply voltages for Optics missing
ON*
1,8 V and 1,4 V for CPU and SDRAM OK
OFF
1,8 V and 1,4 V for CPU and SDRAM
missing
ON*
Supply voltages for PMT OK
OFF
Supply voltages for PMT missing
ON*
Analog supply voltages OK
OFF
Analog supply voltages missing
ON*
CAN Bus enabled
OFF
CAN Bus error
ON
FPGA not configured
OFF*
FPGA configured
ON**
Ethernet traffic
OFF
No Ethernet traffic
ON*
1,2 and 2,5 V for FPGA OK
OFF
1,2 and 2,5 V for FPGA missing
Repair and Service
Chapter 3.4 / Page 5 of 7