Return IP procedure failed.
Meaning: At a handling error situation the IP is not transported into the cassette
successfully.
Reason 1: Handling of the IP not successful.
Cure 1: Restart Digitizer.
Reason 2: Cassette not inserted straight into 'Cassette Unit'.
The IP has been lost at Slow Scan Drum and is placed at IP-Tray.
Cure 2: Check the Cassette for straight entry.
The distances of the arrow marks at tube side of
the inserted Cassette have to be equal to the slot.
In case of repeated appearance of this failure
the 'Cassette Unit' has to be replaced.
Open the Digitizer and remove the lost IP.
Reason 3: IP has jammed.
Cure 3: Open Digitizer and remove jammed IP.
Reason 4: The IP is not transported straight through the Digitizer.
The IP has been lost at Slow Scan Drum and is placed at IP-Tray.
Cure 4: Open the Digitizer and remove the lost IP.
Check by help of Diagnose Cycle the 'Cassette Unit' for
feeding the IP straight into the Digitizer.
In case of repeated appearance of this failure
the 'Cassette Unit' has to be replaced.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 20 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents