AGFA CR Series Service Manual page 131

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2.4.3
Destination Settings
The menu "Destination Settings" allows entering data to modify the destination network
Purpose
settings.
(1)
Select service menu:
Usage
<Destination Settings>.
Perform the required settings. See
also NOTE below.
(2)
Select <Load Destination
Network Data from USB> to
read out the data from an adc.cpf
file on the digitizer USB flash drive.
This CPF file:
(3)
If you want to go back to the
default destination settings: Select
<Load default settings>.
(4)
Select <Test Communication>
to check the network connection to
the NX workstation.
*The CCM tool (Configuration and Customization Manager) is by default installed on the NX workstation.
NOTE:
The configuration of the digitizer settings is described in the digitizer Service
Manual, chapter 1, Controls, Connections and Setup Procedures.
Function "Load default settings" is available as of software version ARC_1404
only.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
Can be created with the CCM
tool* or
Can be exported from the NX
configuration tool.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 15 of 31
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Tools and auxiliary Means
Figure 11
Agfa Company Confidential

Hide quick links: