Image Faults Parallel To Image Plate Run; Calibration Lines - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
After bootup of the digitizer the network settings are as follows:
IP Address:
Subnet Mask:
Gateway IP Address:
Digitizer AE Title:
Digitizer Hostname:
NOTE:
After a reset of the network settings it is required to configure the network settings
again. Refer to Service Manual Chapter 1, "Controls, Connections, and Setup
Procedures".
6

Image Faults parallel to Image Plate Run

6.1

Calibration Lines

Calibration lines are visible in the clinical
Symptom
image or a flatfield image.
Digitizer not calibrated after exchange of optic module or PMT with light collector.
Cause
Perform a shading calibration. Refer to chapter 3.6, Adjustments and Calibrations
Solution
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
192.192.192.192
255.255.255.0
192.192.192.1
DIG_DEFAULT
192.192.192.192
Figure 3: Example for calibration lines
Repair and Service
Troubleshooting
Chapter 3.3 / 13 of 46