Device E-Label data corrupt.
Meaning: The Device E-Label is corrupt.
The identification and serial number is needed for operation.
The Device E-Label is fixed at the mounting rack and cannot be exchanged.
Reason 1: The 'Device E-Label' connection is disturbed.
Cure 1: Check cables and connectors between 'PMI Board' and 'Device E-
Label'.
Replace 'Cable Device E-Label to PMI Board'.
Reason 2: The device eLabel is empty, was formatted or contains invalid data.
Cure 2: Replace Digitizer.
Device E-Label not accessible.
Meaning: The Device E-Label is not detected.
The identification and serial number is needed for operation.
The Device E-Label is fixed at the mounting rack and cannot be exchanged.
Reason 1: The 'Device E-Label' is not connected.
Cure 1: Check cables and connectors between 'PMI Board' and 'Device E-
Label'.
Replace 'Cable Device E-Label to PMI Board'.
Reason 2: The 'PMI Board' is defective.
Cure 2: Replace 'PMI Board'.
Reason 3: The 'Device E-Label' is defective.
Cure 3: Replace Digitizer.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 36 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents