03-2013
Transport Motors stop as Cassette Release Button is pressed during IP
transport.
Meaning: The mechanical Cassette Release Button is pressed while the
Digitizer is still working.
The Motors stop to avoid any damage to IP, Cassette or Digitizer.
Manual intervention needed.
Reason 1: Pressing of the cassette release button before end of cycle.
Cure 1: Please clamp the cassette again by pushing it towards the digitizer.
Restart Digitizer.
Reason 2: 'Cassette Detection Light Sensor' does not work properly.
The Light Sensor is not interrupted by the corresponding lever during IP
transport.
Cure 2: Check Cassette Detection Light Barrier and the corresponding lever.
Check mounting of the 'Cassette Detection Light Sensor'.
Reason 3: 'Cassette Detection Light Sensor' does not work properly.
Cure 3: Check cable and connectors to 'Cassette Detection Light Sensor'.
Replace 'Y-Cable Power Board to Drawer/Cass. LS'.
Reason 4: 'Cassette Detection Light Sensor' defective.
Cure 4: Replace 'Cassette Detection Light Sensor'.
Reason 5: The 'Cassette Unit' has a mechanical defect at the sensor lever.
The Cassette Detection Light Sensor lever is not moving sufficiently towards
the sensor gap.
Cure 5: Check the 'Cassette Unit' for proper mechanical function when a
Cassette is inserted.
Replace 'Cassette Unit'.
Reason 6: 'Power Board' defective.
Cure 6: Replace 'Power Board'.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 22 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents