Edition 1, Revision 2
03-2013
Calculated Image Plate length reached an error limit.
There may be distortions in the scanned image.
Meaning: The scan length changed and is far away from the allowed tolerance
range.
The automatic speed calibration of the Slow Scan drive transmission reached
an error limit.
Reason 1: IP damaged. IP transportation not steady and stable.
This causes a large variation in the correction result from one scan cycle to the
next cycle.
Cure 1: Please check the image plate for damages.
Replace IP.
Reason 2: Slow Scan mechanics damaged. IP transportation not steady and
stable.
This causes a large variation in the correction result from one scan cycle to the
next cycle.
Cure 2: Check belt tension of Slow Scan drum drive. Check the Slow Scan
mechanics for damages.
Check 'Slow Scan Belt'. Check the magnetic drum for damages.
Replace 'Slow Scan Plate with Belt Tensioner'. Replace 'Slow Scan Belt'.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 46 of 46