AGFA CR Series Service Manual page 182

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
CODE
ARC2805
ARC3100
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Check of supply voltages on 'PMI Board' failed.
Meaning: The check of the power supply voltages on the 'PMI Board' failed.
Supply voltage for CAN, Optic, PMT or internal of PMI is missing or not in a
valid range.
Reason 1: Cable defective or disconnected.
Cure 1: Check cable and connectors between 'Power Board' and the 'PMI
Board'.
Replace 'Cable PMI Board to Power Board'.
Reason 2: 'Power Board' is defective.
Cure 2: Replace 'Power Board'.
Reason 3: 'PMI Board' is defective.
Cure 3: Replace 'PMI Board'.
Light inside the digitizer detected.
Meaning: Check for light inside the digitizer failed.
It is too bright to adjust the Photomultiplier.
The optical image sensor is not ready for scan.
Reason 1: Cover not closed.
Cure 1: Close the covers. Restart digitizer.
Reason 2: Ambient Light is too bright.
Cure 2: Protect the digitizer from direct sunlight.
Restart digitizer.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 34 of 46
Agfa Company Confidential