DD+DIS155.11E
Network cable not connected or defective.
Cause 1
Check network interface LEDs at NX workstation: They should light up (constant or
Solution 1
intermittently).
Check network interface LEDs at the digitizer: They should light up (constant or
intermittently).
If the network LEDs on both sides do not light up, try another network cable.
Network cable connected directly between digitizer and workstation and NX
Cause 2
workstation is not configured for a fixed IP address.
Set a fixed IP address for the NX in the "General Settings" window. For instructions
Solution 2
refer to the NX service manual chapter 4, Installation and Configuration. Note, that this
features is available only as of NX software 2.0.8700 / 3.0.8700.
Network interface defective at NX Workstation or digitizer.
Cause 3
Checking the network interface at the NX workstation:
Solution 3
(1)
Connect the Service PC via network cable to the NX workstation (Service PC
IP address configuration not required).
(2)
Check whether the network LEDs at the NX workstation light up.
If the network LEDs on the NX workstation does not light up, exchange the network
interface at the NX workstation.
Checking the network interface at the digitizer:
(3)
Connect the Service PC via network cable to the digitizer (Service PC IP address
configuration not required).
(4)
Check whether the network LEDs at the digitizer light up.
If the network LED on the digitizer does not light up, exchange the PMI board.
Wrong configuration settings on digitizer and/or NX Workstation.
Cause 4
Following settings must be correct on the digitizer and the NX Workstation to allow
Solution 4
communication (see table below):
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Chapter 3.3 / 11 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents