DD+DIS155.11E
(3)
Put a cassette into the digitizer.
(4)
Press the grey marked buttons to
perform the required actions.
See also NOTE below.
IMPORTANT:
The Error Viewer and related confirm / cancel window only pops up at the
NX workstation. This means, that in any case entering the diagnose cycle
window needs to be confirmed at the NX workstation.
Only leave the Diagnose Cycle window by the CANCEL button. Leaving it via
browser back button requires a reboot of the digitizer to allow further service
activities.
NOTE:
Description of the colored rectangles shown in the diagnose cycle overview:
A light blue displayed rectangle shows other statuses of the complete
cassette cycle.
A green displayed rectangle shows the current status.
A dark blue displayed rectangle shows the next possible action.
A gray displayed rectangle shows the next possible action.